Auto Chemistry Analyzer UA300
Feature:
Throughput: Up to 225 tests/Hour(not including ISE tests)
Test Method: End point, Fixed time, Kinetic
Test Sequence: Discrete, Random access, STAT can be inserted at any time.
Spectrometer: Enclosed array of spectrometer and photodiodes, maintenance free.
Specification:
Sample/Reagent/Stirrer system
Sample volume: 2μL~30μL,step by 0.1μL
Sample disc: 60 positions for primary tubes and secondary cups
Reagent volume: 20μL~300μL, step by 1μL
Reagent disc: provide 60 reagent bottle positions
Reagent Cooling: continuous cooling system according to the principle of peltier
Carry over: optimized item test sequence and reinforced washing process, lower the carry over
Reaction system
Cuvette: disposable plastic cuvette
Reaction volume: 150μL~600μL
Temperature: 37℃±0.1℃
Optical system
Light source: 12V 20W, halogen lamp with lifetime more than 2000 hours
Spectrometer: post spectrometer with arrayed and sealed filters, maintenance free
Wavelength: 340nm、405nm、450nm、510nm、546nm、578nm、630nm、670nm
Power/Dimension
Voltage: AC 85V~264V, 55Hz±8Hz, Auto adapt according to the input voltage
Power Consume: ≤350VA
Dimension: 850mm×650mm×1067mm(L×W×H)
Net weight: 90Kg
Write a review
Your Name:Your Review: Note: HTML is not translated!
Enter the code in the box below: